Oxygen Sensor OOM202-1 O2 cell
| Measurement range | 0 % ... 100 % oxygen (at atmospheric pressure) |
| Nominal sensor lifetime | ≥ 1 000 000 % volume oxygen hours |
| Output in ambient air | 13 mV ... 16 mV |
| Electrical interface | 3.5 mm mono phone Jack |
| Accuracy | meets ISO 80601-2-55 requirements |
| Repeatability | < 1 % volume O2 at constant temperature and pressure |
| Linearity error | < 3 % relative |
| Response time | < 12 s to 90 % of final value |
| Zero offset voltage | < 200 µV in 100 % nitrogen, applied for 5 min |
| Cross interference | meets ISO 80601-2-55 requirements |
| Influence of humidity | -0.03 % rel. per % RH at 25 °C |
| Pressure range | 0.6 bar ... 2 bar (ppO2 0 ... 1250 mbar O2) |
| Influence of pressure | proportional to change in oxygen partial pressure |
| Influence of mechanical shock | < 1 % relative after a fall from 1 m |
| Operating temperature | 0 °C ... +50 °C |
| Temperature compensation | built-in NTC compensation |
| Effect of temperature compensation (steadystate) | between +25 °C and +40 °C: 3 % relative error between 0 °C and +50 °C: 8 % relative error |
| Operating humidity | 0 % ... 99 % RH non-condensing |
| Long term output drift | 1 % volume oxygen per month typically < -15 % relative over lifetime |
| Storage temperature | -20 °C ... +50 °C |
| Recommended storage | +5 °C … +15 °C |
| Recommended load | ≥ 10 kOhms |
| Warm-up time | < 30 minutes, after replacement of sensor |
| Weight | approximately 28 grams |
our more items:
1. Reusable/Disposable Spo2 Sensor2. Spo2 extension cable
3. ECG trunk cable &leadwire
4. EKG Cable&Accessories
5. ECG electrodes
6. EEG Leadwires with electrodes
7. IBP Cable
8. Water trap
9. NIBP cuff
10. Temperature Probe
11. Temperature probe interconnect cable
12. Disposable pressure transducer connector
13. Medical oxygen sensor and battery
14. TOCO and US transducer
Hot Tags: oxygen sensor oom202-1 o2 cell, manufacturers, suppliers, factory, cheap, buy