Flow Sensor Solutions in Ventilators

- Sep 30, 2022-

Flow sensors are used to measure the flow rate of blood or oxygen through a vessel. Implantable flow sensors are commonly incorporated into a flexible cuff (Fig. 20.10) that is fitted around the vessel whose flow rate is to be measured.


As the use and spread of ventilators continues to grow so a CMOSens Technology has established a new generation of flow sensors.


Continuous airflow measurements during anaesthesia monitoring, intensive care treatment as well as in clinical and ambulatory environments provide important information for the assessment of cardiorespiratory and breathing circuit behaviour and have become indispensable in modern medicine.

Mechanical ventilation systems supply patients with breathing gas by means of mechanical “air pumps” and this ventilation technique uses positive pressure to deliver air to the patient’s lungs.

Figure 1: Schematic construction of a ventilator with the typical different sensor positions and the use of a humidifier.

The increase of intelligent features incorporated into these ventilators, allow them to automatically adapt to changes in lung function or patient breathing. Modern pressure-controlled or volume-controlled ventilation is therefore now more patient oriented than ever. Since fewer and fewer ventilation modes are required due to the increase in device intelligence, medical ventilators have overall become less complex to operate.

Non-invasive ventilation refers to ventilation therapies that are performed using masks or nasal cannulas. This is often referred to as mask ventilation or NIV/NPPV (non-invasive ventilation or non-invasive positive pressure ventilation). In invasive ventilation, an endotracheal tube or a tracheal cannula is inserted into the trachea of the patient to supply the lungs with air. Both types of ventilation – non-invasive and invasive – have merit and are used in a complementary manner.

A factor that should not be underestimated is the humidification of the inhaled air as it goes far beyond mere patient comfort. Well-humidified and warmed air contributes significantly to the success of ventilation therapy as it improves both secretion drainage and the tolerance of non-invasive ventilation therapy. 

Current trends in hospitals show that non-invasive ventilation is used more frequently today and for far more symptoms than ever before. Intensive care units, for example, increasingly use non-invasive ventilation as a first-line of treatment, which reduces infectious complications, weaning periods, ICU stay lengths, intubation rates and costs.

The key issue for all ventilators is the accurate measurement of the breathing gas flow rate and the volume of breathing gas that flows in and out of the patient. These measurements with the highest sensitivity and accuracy enable the previously mentioned and nowadays prevailing patient-oriented ventilation, which also better reflects the pathophysiology of the patient. Figure 1 shows the schematic construction of a ventilator with the typical air flow/sensor positions.

Technical challenges

The complex breathing circuits have a wide range of composition variability due to the different types of tubing, humidifiers, filters and adapters used. This often results in leaks and imperfections, which is why the inspiratory flow rate (I) sometimes differs significantly from the flow rate that actually reaches the patient. The same applies to the expiratory flow rate (E). Airflow measurements are also hampered by the constant changes in air temperature, humidity and breathing gas composition as well as the contamination of hoses and expiratory/proximal sensors with sputum, pathogens and blood. Due to technical limitations, measurements of inspiratory (I) and expiratory flow rates (E) were performed inside the ventilator in the past. The rough flow values were then corrected as far as possible using complex and often inaccurate compensation algorithms.



Figure 2.    Schematic of a ventilation setup with extremely humid air and a very small tidal volume of only 5 ml. 

 

Proximal flow sensors must be reliable and cost effective, long-term stable and, moreover, feature numerous other ventilator-specific features in order to be suited for modern patient-oriented ventilation.    In addition, particularly stringent requirements for hygienic sterilization are needed as the sensors come into contact with air that is potentially contaminated with pathogens. 

 

The Achilles’ heel of all current air flow sensors is the use in combination with humidifiers.    High humidity becomes a problem when it leads to condensation, causing macroscopic water droplets to rain out in in the cooler parts of the ventilator circuit.    As a solution, all Sensirion proximal and expiratory sensors are equipped with an additional external heating element.    Operation of this heating element with a maximum of 0.5 W is sufficient to reliably prevent condensation in the sensor and thus ensure long-term stable and reliable operation. 

 

The schematic illustrated in Figure 2 shows a humidifier typically used in ventilator setups to ensure the breathing air is well humidified.    The steel cylinder in the oven is kept at 37°C and simulates the lungs with the connected pressure sensor used as a reference.    The controlled valve is closed during the inspiratory breathing cycle and opened once per second for the expiratory part of the breathing cycle. 

 

Without the use of the heater, individual drops of water can run over the sensor element and cause misreading of the measurement values.    This misreading can clearly be recognized by the deviations of the expiratory/inspiratory volume from the reference volume. 

 

Outlook 

 

The use and spread of ventilators will continue to grow strongly in the future due to the increasing number of lung diseases.    Modern ventilators place ever-growing demands on sensors in order to place the focus on the patients and their therapy. 

 

The CMOSens Technology has established a new generation of flow sensors that have proven their reliability millions of times in the field of CPAP devices and automotive applications with the advantages for ventilators being evident. 

 

It is the technological advantage that will enable manufacturers to realise the next quantum leaps in ventilation.