Electrode Placement 3 Leads

- Aug 14, 2020-

Electrode Placement (3 Leads)


         Electrode Placement (3 Lead) Right intra-clavicular fossa R Red  Left intra-clavicular fossa. L Yellow  Between 6th and 7th intercostal space on the left midclavicular line N Black  Lead connections (3 Lead)  Steps for application of ECG electrodes :  Proper skin preparation is necessary for the good quality signal to pick up and display. Please follow the guidelines listed below:  Wash electrode site and shave surface hair.  Gently rub skin surface with a prep pad to remove outer epidermal layer  Thoroughly clean the site with soap and water, depending on your patient’s skin type and sensitivity.